Otsikko Sterilization of health care products — Chemical indicators — Part 4: Type 2 indicator systems as an alternative to the Bowie and Dick-type test for the detection of steam penetration
Otsikko (en) Sterilization of health care products — Chemical indicators — Part 4: Type 2 indicator systems as an alternative to the Bowie and Dick-type test for the detection of steam penetration
Tunnus ISO/DIS 11140-4
Julkaisija ISO
Komitea ISO/TC 198 (Sterilization of health care products)
Lausuntokierroksen aloituspäivämäärä
Lausuntopyynnön määräaika
Toimialayhteisö

Ehdotuksen soveltamisala

This document specifies the performance for a Type 2 indicator system to be used as an alternative to the Bowie and Dick-type test for steam sterilizers for wrapped health care goods (instruments, etc., and porous loads). An indicator system conforming with this document incorporates an indicator together with a specified material which is used as a test load. This test load may, or may not, be re-usable. This document does not specify requirements for the test load, but specifies the performance of the indicator system (the indicator in combination with the test load with which it is intended to be used). The indicator system specified in this document is intended to identify poor steam penetration but does not necessarily indicate the cause of this poor steam penetration. This document does not include test methods to establish the suitability of these indicator systems for use in sterilizers in which the air removal stage does not include evacuation below atmospheric pressure.

Instructions

Rekisteröidyttyäsi Lausuntopyyntöpalvelun käyttäjäksi voit lukea standardiehdotuksia ja kommentoida niiden sisältöä.

Kun haluat kommentoida lukemaasi tekstiä, valitse Lue ehdotus ja anna muutosehdotuksesi kohdassa Jätä kommentti.

Voit myös kertoa mielipiteesi standardiehdotuksesta kokonaisuutena. Ota kantaa siihen, hyväksytkö standardiehdotuksen vahvistamisen standardiksi.